5.3.17 Recorder module
1. Press the Record Wave module key and check that the module starts recording the
selected waveforms. Press the
Stop module key to stop recording.
2. Check that the quality of the recordings is acceptable.
"
5.3.18 Network connection
1. Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the CPU/UPI4NET Board.
Check that the monitor connects to the network, i.e. the network symbol appears under
the clock on the upper right-hand corner of the screen. Also a message regarding the
connected Central should appear in the message field on the screen.
"
5.3.19 Interface Module, E-INT
1. Make sure that the monitor receives all necessary parameter data from the connected
devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup – Interfacing
"
5.3.20 Interface module for PSM, E-INTPSM
1. Make sure that the monitor receives all necessary parameter data from the connected
E-PSM(P) module. Check the screen configuration and the related interfacing settings, if
necessary.
Monitor Setup – Interfacing
"
5.3.21 Device Interfacing Solution, N-DISxxx
1. Make sure that the monitor receives all necessary parameter data from the connected
devices. Check the screen configuration and the related interfacing settings, if necessary.
Check also via the Interfacing menu that the connected DIS module status is correct:
Monitor Setup - Interfacing - Status Page
"