3.3 Central Unit; S/5 8-Module Frame, F-CU8
Figure 7 Central Unit: S/5 8-Module Frame, F-CU8
The front of the Central Unit, F-CU8, houses plug-in parameter modules which interface the
system with external devices. The back of the Central Unit houses the system circuit boards,
together with expansion slots which accommodate PC boards which interface with external
devices and the Datex-Ohmeda Network.
3.3.1 Connecting to mains
Connect the power cord to the mains power inlet on the rear of the Central Unit and to the wall
socket.
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle.
3.3.2 Connecting to the Datex-Ohmeda Network
To connect the monitor to the Datex-Ohmeda Network, make sure a CPU Board, B-CPU6 is
installed.
Use the Monitor-Network cable to connect the monitor to the network as follows:
1. Make sure that the power to the Central Unit is turned off.
2. Connect the RJ-45 network connector to the connector X10 on the CPU board, B-CPU6
and the Identification Plug to the corresponding connector on the B-UPI4NET.
3. Connect the other RJ-45 connector to the corresponding connector on the wall box.
4. Turn on the power to the Central Unit. Enter the Network service menu:
Monitor Setup - Install/Service (password 16-4-34) -
Service (password 26-23-8) -
Frame – Network
5. Make sure that the monitor's network communication has been set according to the used
network software:
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
If necessary, change the monitor's network communication by selecting DRI Level and
turning the ComWheel.
NOTE: If the DRI level is changed, the monitor will restart automatically.