Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
66
Document no. M1137263-04
− F-CU8: the equipotential tap and all the screws are tightened properly.
− sampling line is connected to the Airway Module, if installed.
Check that modules go in smoothly and lock up properly in all module slots.
CAUTION Ensure that the module is properly orientated (i.e. module release latch facing
downward) before insertion.
"
5.3 Functional inspection
WARNING Handle the water trap and its contents as you would any body fluid.
Infectous hazard may be present.
5.3.1 General
1. F-CU5: Connect all the parts together. Check that the cables and screws are tightened
properly. Connect the mains power cord to the N-AC Power Unit.
F-CU8: Connect the mains power cord to the F-CU8.
Check that the stand-by LED is lit.
2. Switch the monitor on.
Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the
loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears.
No error messages should appear on the screen.
3. Configure the screen for the parameters that are connected.
4. Enter the Service Menu.
When applicable, check from the corresponding Parameters submenu that the Timeouts,
Bad checksums and Bad c-s by mod values of inserted modules are not increasing faster
than by 5 per second. Check also that the module memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all state OK.
If connected, the recorder should record two lines of start-up information.
"
Preset the measurement settings for those parameters that are connected, for example:
Record/Print - Record Waveforms - Waveform 1 - ECG1
- Waveform 2 - P1
- Waveform 3 - P2
Invasive Pressures - P1 ‘ART’ Setup -- Label – ART
- P2 ‘CVP’ Setup -- Label – CVP
- P3 Setup -Label –PA
- P4 Setup -Label -P4
- P5 Setup -Label -P5
- P6 Setup -Label -P6
Pulse Oximetry -Pleth Scale –AUTO
or
Others -SPO2 Setup - Pleth Scale –AUTO