3 Service procedures
3.1 General service information
Field repair is limited to replacing parts that are listed in the corresponding spare part lists.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation and a detailed description of the fault.
WARNING Only trained personnel with appropriate equipment should perform the
tests and repairs outlined in this section. Unauthorized service may void
warranty of the unit.
3.1.1 LCD Display, D-LCC12A
The D-LCC12A contains a replaceable backlight.
3.1.2 15” Flat Panel Display, D-FPD15
Maintenance is limited to the basic adjustments through the display's Control dial.
The display should be returned to GE Healthcare for repair.
3.1.3 19” Medical Grade Display
Maintenance is limited to the basic adjustments through the display’s Control dial.
The display should be returned to GE Healthcare for repair.
3.2 Service check
These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include check forms, which may be used when performing the procedures.
The symbol
"in the instructions indicates that the check form contains space to record the
results of the particular procedure.