3.2.4 General inspection
15. Communication lines
After any service repair for the Module mother board, measure resistance from the
following Module mother board’s connector pins against the ground:
Check that the resistance on each of the pins is higher than 10 kΩ. If not, replace the
Module mother board.
"
• Reconnect the PC boards and secure them to the Central Unit with screws.
NOTE: Make sure that the cables are properly attached and secured.
• Install the display together with the Command Bar, the Hemodynamic module and E-REC.
Connect the mains power cord and turn the monitor on.
16. Fan
Check that the fan is running.
"
17. Starting
Check that the monitor starts up properly, i.e. the alarm LEDs on the Command Bar turn
blank, the start-up sound is heard from the loudspeaker and the normal monitoring
screen appears. No error messages should appear on the screen.
"
18. Module communication
Check that the connected modules are recognized, i.e. the needed parameter information
is shown on the screen and the E-REC records two lines of start-up information.
If some parameter information is missing, check the screen configuration from the
MONITOR SETUP menu.
NOTE: InvBP waveforms are not shown without a patient simulator.
"
Module mother board connector Pin 1 +Reset RS485
Pin 5 -Data RS485
Pin 6 +Data RS485
Pin 8 -Reset RS485
Pin 13 Ground