Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
70
Document no. M1137263-04
Perform a NIBP measurement and check that the module gives a reasonable measured
result.
"
5.3.9 Pressure/Pressure Temp Modules, E-P, E-PT
Invasive blood pressure measurement
1. Check the function of the front panel Zero P3 key.
2. Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
Temperature measurement
3. Check the temperature channels with a patient simulator.
Check that the temperature measurement information is shown on the monitor screen as
configured on the simulator.
"
5.3.10 Dual pressure Module, E-PP
1. Check the function of the front panel Zero P5 and Zero P6 keys.
2. Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
"
5.3.11 Cardiac Output Modules, E-COP, E-COPSv
Invasive blood pressure measurement
1. Check the function of the front panel Zero P4 key.
2. Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
Cardiac Output measurement
3. Check the function of the front panel Start C.O. key.
"