S/5
TM
 UPINET Board, B-UPI4NET-02
5
Document no. M1137269-02
2.2.2 RS232 Serial data interface
RS232 Serial data connector X8
2.2.3 DIS interface (RS422)
DIS connector, X5
2.2.4 Synchronization interface  
Synchronization connector, connector X7 (B-UPI4NET -02)
Defibrillation Sync (pin 1)
Defibrillation Sync indication is generated by ECG. When active, the signal is in state 1. After 10 
ms the signal is reset to state 0. New Defibrillation Sync is not generated before the indication is 
deactivated. The delay from the R wave peak to the start of the signal is maximally 35 ms.
9 pin male D-connector Pin Signal
1
2
3
4
5
6
7
8
9
GND
RxD
TxD
NC
GND
N/C
RTS
CTS
N/C
10 pin female connector Pin Signal
1
2
3
4
5
6
7
8
DIS_out +
DIS_out -
15V_DIS
GND
8V_DIS
GND
DIS_in +
DIS_in -
Mini DIN7 connector Pin Signal
1
2
3
5
6
7
8
Defib_sync_out
Reserved
Analog GND
Digital GND
GND
Pressure_out
Direct_ECG_out
6
9
1
5