Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
38
Document no. M1137263-04
3. Connect the AC adapter’s DC plug into the displays power input. Connect the provided
power cord in to the inlet of the auxiliary AC adapter. Then, connect plug to the mains
outlet.
4. Connect the DVI cable to the connector X2 on the CPU board, B-CPU6. Tighten the finger
screws.
A medical grade power supply adapter and the country-specific power cord are included in the
display delivery.
The following cables are included: Analog signal cable (1.8 m long), DVI signal cable (1.8 m long).
Adjustments
The optimal display resolution is SXGA (1280 x 1024). To use the display, a CPU board, B-CPU6
or Display Controller Board B-DISPX is needed.
Changing the screen resolution: Press
Monitor Setup,
select Install/Service (password 16 - 4 - 34),
select Installation - Display Setup.
In Screen 1 ... Screen 3, select the resolution SXGA (1280 x 1024).
After changing the resolution turn the monitor off and on again.
For details about the general use of the 19" Medical Grade Display refer to the display
manufacturer's user guide delivered with the display.
3.6 Display controller boards
CPU board, B-CPU6 has an integrated display controller, which acts as a controller for the
primary display. In case only one display is used, an optional display controller board is not
needed. Optional displays must be supported by different separate display controller boards,
i.e. the primary display by a primary display controller integrated on the CPU board, B-CPU6,
the secondary display by an optional display controller board and the third display by an
optional secondary display controller board. Consequently, it is possible to configure the
displays separately. One of the displays can, for instance, be used for a continuous display of
trend pages.
NOTE: It is possible to use one display controller to obtain duplicate video signals for two
displays by using the display Y-cable.
3.6.1 Jumper settings
The optional display controller boards require different address settings to operate correctly.
The settings are made through dip switches on B-DISPX. Different settings are illustrated in the
following figures.
Figure 17 Address dip switch settings, B-DISPX