Functional check
65
Document no. M1137263-04
see section “Adjustments and calibrations” in Compact Airway modules slot in E-Modules
Technical Reference Manual.
For details on recommended accessories see “Supplies and Accessories” catalog.
5.1.1 Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic
module, and the accessories needed:
Table 14 Patient simulators’ compatibility with each hemodynamic module
Table 15 Adapter cables for hemodynamic patient simulators
5.2 Visual inspection
Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor.
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable
from the monitor. If the Memory Module, E-MEM is connected, remove any memory cards.
1. Check all units visually
Check that all parts are intact and that the cables and screws are connected and
tightened properly. Especially check the following parts:
Video displays: the display power cord is locked to the display.
F-CU5(P), F-CPU and N-AC: all the screws and the equipotential tap on the N-AC are
tightened properly.
Module Parameter
Patient simulator
M1010831 MedSim 874027
E-PSM(P), E-PRESTN ECG Ok Multilink ECG acc. Not compatible
T 402015-004 402015-004 and
M1010832
InvBP Ok M1010858
Patient simulator Adapter cables for simulators
Hemodynamic patient simulator - Dual temperature adapter cable 402015-004
Hemodynamic patient simulator - Dual Inv.BP adapter cable 2005772-001
Medsim - Temperature adapter cable M1010832
Medsim - Inv.BP adapter cable M1010858
Lionheart & MPS450 - Temperature adapter cable M1010846
Lionheart & MPS450 - Inv.BP adapter cable M1010862
ECG