Interfacing
57
Document no. M1137263-04
Table 8 Events transferred from S/5 Anesthesia Delivery Unit to the
S/5 Anesthesia Monitor
4.4 Interfacing Dräger Cicero, Cato, Julian and
Narkomed 2C (by NAD)
It is possible to interface Dräger Cicero and Cato to the S/5 Anesthesia Monitor or to the
S/5 Critical Care Monitor via the Interface Module, E-INT.
NOTE: The Interface Board must be equipped with revision 881652-4.0 software or later.
The parameters transferred to the S/5 Anesthesia Monitor or S/5 Critical Care Monitor are
summarized in Table 9 and Table 10.
NOTE: Some of the summarized parameters (waveforms or numerics) may not be available in
all device versions.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Events transferred once a minute
Ventilator settings Ventilation (mode)
Tidal Volume
Minute Volume
Resp. Rate
I:E Times: inp
I:E Times: exp
InspPause
Ventilator measurements Ppeak
Pplat
Peep
Pmin
Fresh gas settings Agent name
Agent % in fresh gas
Total flow
O
2
flow
N
2
O flow
Air flow
Events transferred during system check
System check test results Gas Delivery: Agent
Gas Delivery: N
2
O
Leak Tests: AUTO
AUTO Leak (ml/min)
Leak Test: MAN
MAN Leak (ml/min)
Number of confirmed checklist items
Bypass Check