Interfacing
59
Document no. M1137263-04
Table 10 Parameters transferred from Dräger Cato, Julian and Narkomed 2C
(NAD) monitor to the S/5Anesthesia Monitor or S/5Critical Care
Monitor
NOTE: The transference of parameters illustrated in Table 7 and Table 8 represents the
maximum. Some of these parameters may not be available in all device versions.
NOTE: The transference of waveforms requires the MEDIBUS real-time extension and, in Cato
monitors, waveforms are available only on port 1.
Selection Waveforms Numerics States
Gases / SSS CO
2
(kPa, mmHg, %) CO
2
Fi & Et (kPa, %)
O
2
Fi & Et(%)
N
2
O Fi & Et (%)
RR (CO
2
) (1/min)
AA Fi & Et (%)
Tidal Volume insp. (1)
Minute Volume exp. (1)
Ppeak (mbar)
Pplat (mbar); not in Narkomed 2C
PEEP (mbar)
None
SpO
2
Pleth (% full scale) SpO
2
(%)
Pulse rate (1/min)
Probe Off