Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
20
Document no. M1137272-02
NOTE: After changing the resolution, the monitor must be restarted before the setting becomes
active.
"
5. Brightness and contrast adjustments
Press the Control dial while the LCD display is on to activate the OSD menu
(on-screen-display).
NOTE: Pressing the Control dial for more than 2 seconds turns the power off.
Perform the contrast and brightness adjustments if needed. Please refer to the display's
User's Manual.
6. Color temperature adjustments
Perform the color adjustments if needed. Please refer to the display's User's Manual.
7. Auto Setup
Perform Auto Setup to automatically adjust the screen size, position, phase and clock, if
needed.
NOTE: This adjustment is available only with analog input (VGA cable). This function is not
available for digital input (DVI).
NOTE: Position adjustments, clock and phase can be adjusted also manually. Please refer to the
display's User's Manual.
Exit the OSD menu and go back to the normal screen.
"
8. Electrical safety check
Perform an electrical safety check and a leakage current test.
"
9. Functioning after safety check
Check that the LCD display functions normally after the performed electrical safety check.
"
10. Final cleaning
Clean the LCD display with suitable detergent.
"
Fill in all necessary documents.