Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
A-2(2)
Document no. M1125635-05
OK N.A. Fail OK N.A. Fail
16. Fan
17. Starting
18. Module
communication
19. Real time clock
20. Loudspeaker sound
21. Monitor software L-
22. Content of service log
23. Voltages
+15 V 14.20...15.60 V
-15 V -14.00...-15.50 V
+15 VD 14.10...15.60 V
+5 V 4.70...5.40 V
OK N.A. Fail OK N.A. Fail
24. Watchdog circuitry
25. Trend retaining
26. Service reset button
27. Recovering from
power loss
Notes
28. Electrical safety check
29. Functioning after
electrical safety check
30. Final cleaning
Notes
Used spare parts
Signature