Document no. M1137263-04
Cautions
CAUTION A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Installation
• Leave space for air circulation to prevent the monitor from overheating.
• Ensure that the module is properly orientated (i.e. module release latch facing downward)
before insertion.
• Before connecting the power cord to the power supply, check that the local voltage and
frequency correspond with the rating stated on the device plate. See instructions for
different displays in section “Displays”.
• Turn off the power before making any rear panel connections.
Before use
• Allow two minutes for warm-up and note any error messages or deviations from normal
operation.
• Clean the rear panel fan dust filters once a month or whenever necessary.
• Do not connect a sampling line to the female Patient Spirometry connector while the
other end of the sampling line is connected to the D-fend water trap. The pressure in the
gas sampling system may cause damage to the PVX unit pressure transducers.
Fuse replacement
• Replace a fuse only with one of the same type and rating.
Cleaning and service
• Do not use hypochlorite-, acetone-, phenol- or ammonia -based cleaners, abrasive
material or harsh chemicals as they may damage the surfaces of the device.
• Do not use abrasive cleaning compounds, instruments, brushes or rough-surface
materials.
• Do not apply pressurized air to any outlet or tubing connected to the monitor. Pressure
may destroy sensitive elements.
• Do not clean the spirometry tubes with high pressure air or O
2
flushing while the
spirometry tubes are connected to Patient Spirometry connector. High differential
pressure may damage PVX unit pressure transducers.
Special components
Storage and transport
Do not store or transport the monitor outside the specified temperature, pressure and humidity
ranges:
Temperature -10...+50 °C/14...122 °F
Atmospheric pressure 660...1060 hPa/500...800 mmHg/660...1060 mbar
Relative humidity 10...90% noncondensing
For display specific environmental requirements see specifications in the “Display” slot.
• A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery
according to local regulations.