Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
10
Document no. M1137269-02
7. Watchdog
Test the UPI4NET board watchdog function.
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
- Set/Test – WD by UPI
Perform the test and check that monitoring continues normally.
"
Net function
8. Recognition of ID-plug
Ensure that the monitor is connected to the network. Switch the monitor to standby.
Disconnect the Identification plug from the UPI4NET board. Switch the monitor back on
and check that the message “Check network connectors” appears in the message field.
Reconnect the Identification plug and check that the monitor connects to the network.
"
General
9. Electrical safety check
Perform an electrical safety check and a leakage current test.
"
10. Functioning after electrical safety test
Check that the system functions normally after the performed electrical safety check.
"
• Fill in all necessary documents.