10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MEASure:WITHstand:CURRent?
Response
syntax
<data>
<data>
Withstand-voltage test measured current value (NR1 or NR2)
Function
Queries the measured current value (unit: mA) for withstand-voltage tests.
Example
Queries the measured current value for withstand-voltage tests.
Transmission :MEASure:WITHstand:CURRent?
Response 2.00
The measured current value is 2.00 mA.
Error
The execution of this command in a state other than the TEST state causes an execution error.
Syntax
:MEASure:WITHstand:TIMer?
Response
syntax
<data1>, <data2>
<data1>
Withstand-voltage test time elapsed (NR1 or NR2)
<data2>
Timer types:
0
(test time),
1
(ramp-up time),
2
(ramp-down time) (NR1)
Function
Queries the test time elapsed (unit: s) for withstand-voltage tests.
Example
Queries the test time elapsed for withstand-voltage tests.
Transmission :MEASure:WITHstand:TIMer?
Response 30.0,0
The elapsed test time is 30.0 s, and the type of timer is test time.
Error
The execution of this command in a state other than the TEST state causes an execution error.
:MEASure:WITHstand:CURRent?
Queries the measured current value for withstand-voltage tests.
:MEASure:WITHstand:TIMer?
Queries the test time elapsed for withstand-voltage tests.