186
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MEMory:WITHstand:SAVE <data>
<data>
1to10
(NR1)
Function
Saves the current withstand-voltage test settings in the memory file specified in <data>.
Example
Saves the settings for withstand-voltage tests in memory file 2.
Transmission :MEMory:WITHstand:SAVE 2
Error
The execution of this command in a state other than the withstand-voltage mode READY state
causes an execution error.
The execution of this command in the insulation-resistance mode and program mode also causes
an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
:MEMory:WITHstand:CLEar <data>
<data>
1to10
(NR1)
Function
Clears and resets the data saved for the withstand-voltage test specified in <data>.
Example
Clears and resets the contents of memory file 3 for withstand-voltage tests.
Transmission :MEMory:WITHstand:CLEar 3
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
:CONFigure:INSulation:VOLTage <data>
<data>
50 to 1200
(NR1)
Function
Sets the test-voltage value (unit: V) for insulation-resistance tests, in the READY state.
Example
Sets the test-voltage value for the insulation-resistance tests to 500 V.
Transmission :CONFigure:INSulation:VOLTage 500
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
The lower-limit value restricts the test-voltage-value setting range (see Section 4.3.1 Setting of
Test-Voltage Values). The transmission of a value outside the range causes an execution error.
:MEMor
:WITHstand:SAVE
Saves set values for withstand-voltage tests in memory.
:MEMor
:WITHstand:CLEar
Clears the set-value memory for withstand-voltage tests.
:CONFi
ure:INSulation:VOLTa
e
Sets the test-voltage value for insulation-resistance tests.