10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:INSulation:VOLTage?
Response
syntax
<data>
<data>
50 to 1200
(NR1)
Function
Queries the test-voltage value (unit: V) for insulation-resistance tests.
Example
Queries the test-voltage value for insulation-resistance tests.
Transmission :CONFigure:INSulation:VOLTage?
Response 500
The test-voltage value is 500 V.
Syntax
:CONFigure:INSulation:RUPPer <data>
<data>
0.10 to 9999
(NR1)
Function
Sets the upper-limit value (unit: MΩ) for insulation-resistance tests, in the READY state.
Example
Sets the upper-limit value value for the insulation-resistance tests to 2000 MΩ.
Transmission :CONFigure:INSulation:RUPPer 2000
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:CONFigure:INSulation:RUPPer?
Response
syntax
<data>
<data>
0.10 to 9999
(NR1 or NR2)
Function
Queries the upper-limit value (unit: MΩ) for insulation-resistance tests.
Example
Queries the upper-limit value for insulation-resistance tests.
Transmission :CONFigure:INSulation:RUPPer?
Response 2000
The upper-limit value is 2000 MΩ.
:CONFi
ure:INSulation:VOLTa
e?
Queries the test-voltage value for insulation-resistance tests.
:CONFi
ure:INSulation:RUPPer
Sets the upper-limit value for insulation-resistance tests.
:CONFi
ure:INSulation:RUPPer?
Queries the upper-limit value for insulation-resistance tests.