10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:CONFigure:INSulation:RLOWer <data>
<data>
0.10 to 9999
(NRf)
Function
Sets the lower-limit value (unit: MΩ) for insulation-resistance tests, in the READY state.
Example
Sets the lower-limit value value for the insulation-resistance tests to 10.0 MΩ.
Transmission :CONFigure:INSulation:RLOWer 10.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
The test-voltage value restricts the lower-limit-value setting range (see Section 4.3.2 Setting of
Lower-Limit Values (Upper-Limit Values)). The transmission of a value outside the range causes
an execution error.
Syntax
:CONFigure:INSulation:RLOWer?
Response
syntax
<data>
<data>
0.10 to 9999
(NR1 or NR2)
Function
Queries the lower-limit value (unit: MΩ) for insulation-resistance tests.
Example
Queries the lower-limit value for insulation-resistance tests.
Transmission :CONFigure:INSulation:RLOWer?
Response 10.0
The lower-limit value is 10.0 MΩ.
Syntax
:CONFigure:INSulation:TIMer <data>
<data>
0.3 to 999
(NRf)
Function
Sets the test time (unit: s) for insulation-resistance tests, in the READY state.
Example
Sets the test time for the insulation-resistance tests to 10.0 s.
Transmission :CONFigure:INSulation:TIMer 10.0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
:CONFi
ure:INSulation:RLOWer
Sets the lower-limit value for insulation-resistance tests.
:CONFi
ure:INSulation:RLOWer?
Queries the lower-limit value for insulation-resistance tests.
:CONFi
ure:INSulation:TIMer
Sets the test time for insulation-resistance tests.