191
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:INSulation:TIMer?
Response
syntax
<data>
<data>
ON/OFF
(Character data)
Function
Queries the test time enablement for insulation-resistance tests.
Example
Queries the test time enablement for insulation-resistance tests.
Transmission :INSulation:TIMer?
Response ON
Syntax
:INSulation:DELay <data>
<data>
ON/OFF
(Character data)
Function
Sets the delay time ON/OFF for insulation-resistance tests in the READY state.
Example
Sets the delay time for insulation-resistance tests to ON.
Transmission :INSulation:DELay ON
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:INSulation:DELay?
Response
syntax
<data>
<data>
ON/OFF
(Character data)
Function
Queries the delay time ON/OFF for insulation-resistance tests.
Example
Queries the delay time ON/OFF for insulation-resistance tests.
Transmission :INSulation:DELay?
Response ON
:INSulation:TIMer?
Queries the test time enablement for insulation-resistance tests.
:INSulation:DELa
y
Sets the delay time ON/OFF for insulation-resistance tests.
:INSulation:DELa
y
?
Queries the delay time ON/OFF for insulation-resistance tests.