10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MEASure:INSulation:RESistance?
Response
syntax
<data>
<data>
Insulation-resistance test measured resistance value (NR1 or NR2)
O.F.: 10000
U.F.: 0
Function
Queries the measured resistance value (unit: MΩ) for insulation-resistance tests.
Example
Queries the measured resistance value for insulation-resistance tests.
Transmission :MEASure:INSulation:RESistance?
Response 100.0
The measured resistance value is 100 MΩ.
Error
The execution of this command in a state other than the TEST state causes an execution error.
Syntax
:MEASure:INSulation:TIMer?
Response
syntax
<data>
<data>
Insulation-resistance test time elapsed (NR1 or NR2)
Function
Queries the test time elapsed (unit: s) for insulation-resistance tests.
Example
Queries the test time elapsed for insulation-resistance tests.
Transmission :MEASure:INSulation:TIMer?
Response 10.0
The test time elapsed is 10.0 s.
Error
The execution of this command in a state other than the TEST state causes an execution error.
:MEASure:INSulation:RESistance?
Queries the measured resistance value for insulation-resistance tests.
:MEASure:INSulation:TIMer?
Queries the test time elapsed for insulation-resistance tests.