Model 831 Manual Manually Controlling the Internal Noise Source 15-19
If the trigger source is set to Z or
Mid Band, a cursor will not be
shown.
FIGURE 15-27 Linked Controls
Hints on Making a Good RT-60 Measurement
Make sure the exit time is long enough to allow egress
and any residual noise to die down
Turn off your cell phone and remove other possible
sources of sound contamination.
Be quiet during the background level measurement
(between the exit period and filling the pre-trigger buffer.
See the “Measurement State Sequence” on page 15-5’.
Minimize the ambient noise in the room by turning off
the mechanical equipment (i.e., HVAC units).
Close doors, windows, and partitions to reduce
reverberant energy from adjoining areas.
Use a sound source with repeatable amplitude and
spectral content.
Use the exclude feature when something goes wrong. It
is a good idea to configure the instrument to take a few
more decays than what is required just in case.
Trigger near the maximum level. For interrupted noise
method measurements, a few trial runs may be required
to determine a good trigger point due to the requirement
that the noise source level be above the trigger level for a
specific duration (Build Time).