15-20 Customizing Measurements Model 831 Manual
Energize the room sufficiently. It may take multiple
sound sources and measurements to characterize the
decay times over the full desired spectrum.
Make measurements at many different source-
microphone combinations to improve the ensemble
uncertainty. See the “Accuracy Grade section on page
13-28 for more information.
Minimize what is in the room (see ISO 3382-2:2008(E)
Section 4.1).
ASTM E2235 requires working with a diffuse field
(random incidence) microphone. You can select the
correction FF->RI to adapt your microphone
characteristics if needed.
FIGURE 15-28 Microphone Correction
Customizing Measurements
From any of the pages on the RT-60 tab, press the Center
Softkey labeled Menu, highlight Settings, press
5 to
display the settings tabs shown below. These pages are used
to modify the measurement parameters. The active
parameter values shown represent the default values.