Model 831 Manual Live Tab 5-9
FIGURE 5-9 426A12 Preamp, E.A. On
In FIGURE 5-9 we see the display includes the serial
number and the current internal temperature, humidity and
dew point. The temperature, humidity and dew point metrics
permit desiccant maintenance information to be gathered
regularly and stored with the data. This can also be done
remotely without having to perform an in-field service call.
The electrostatic actuator can be activated when stopped by
pressing
5 to toggle the E.A. calibrator On and Off. The
SPL displays the measured E.A. level.