V Series Operating Instructions 9 - 35
ECG – Arrhythmia 12 Lead Interpretation
5. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
9.9.1.4 Configuring 12 Lead Advanced Setup
1. Select the Procedures tab from the Navigation Area.
2. Select the 12 Lead button.
3. Select the Advanced tab (as shown in FIGURE 9-28).
FIGURE 9-28 Example 12 Lead Dialog (Advanced Tab)
4. Select the buttons described in TABLE 9-7 to configure the tab.
NOTE: This button is unavailable when the Lead Size
button is configured to 0.5 cm/mV.
Tachycardia Configure the 12 Lead tachycardia
threshold.
80 BPM to 130 BPM
The factory default value is 100 bpm.
Bradycardia Configure the 12 Lead Bradycardia
threshold.
40 BPM to 60 BPM
The factory default value is 50 bpm.
V3 Placement Configure the V3 placement Standard and V4R
Interpretation Configure whether or not to display
critical value interpretation results on
the 12 lead report.
On and Off
TABLE 9-6
BUTTON DESCRIPTION CHOICES*
*The factory default configuration settings are in bold.