nP 084 appears on the display.
25. Press OK.
CLIEnt (VNC client) appears on the display.
26. Press OK.
The more signicant part of the address* 255.255. (example) appears on the
display.
27. Press the cursor keys to select and set the desired numbers.
28. Once the last number is selected and set, press to set the less signicant part of
the mask*.
255.255. (example) appears on the display.
29. Press the cursor keys to select and set the desired numbers.
30. Press OK.
nP 085 appears on the display.
31. Press OK.
EtAddr (hardware address/MAC ID) appears on the display.
32. Press OK.
The more signicant part of the address* 00.90.6C. (example) appears on
the display.
33. Press to display the less signicant part of the mask*.
70.49.45. (example) appears on the display.
34. Press OK.
SAVE appears on the display.
35. Press OK.
YES appears on the display.
36. Press OK.
SAVE ... appears on the display during the save process.
37. Press "Exit" to exit the Setup menu.
7 Commissioning X3 Process Indicator PR 5410
EN-141 Minebea Intec