EasyManuals Logo

Minebea Intec X3 Process Indicator PR 5410 User Manual

Minebea Intec X3 Process Indicator PR 5410
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
[Standstill range]
As long as the weight uctuations remain within this range, the device is determined to
be stable.
The input for the parameter [Standstill range] is expressed in "d."
Valid range: 0.0110.00 dc.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect lter setting, resolution too high, standstill condition too
strict).
The input is done as seconds.
Valid range: 0.0<2.5>25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Dene a ±range around the zero point determined by the dead load during calibration;
within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or by
a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.0010000.00 d
For use in legal metrology a value ≤2% of the Max must be entered, example: 60 d for
3000 e of class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations.
Setting range: 0.2510000.00 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any
more.
Setting range for automatic tracking increments: 0.2510 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.125 s
At 0.0 s the zero tracking is switched o.
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum load (Max) without error message.
Setting range: 09999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Minimum weight]
Minimum weight at which a print command can be triggered.
Setting range: 09999999 d
For use in legal metrology a value of at least 20 d must be set.
7 Commissioning X3 Process Indicator PR 5410
EN-175 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish