Specications
Description Data
Connection D-Sub female connector, 15-pin
Output: Qty 1 active current output:
0/420 mA (max. 24 mA), 10 V output voltage via external
500 resistor
Output: Function Gross/net weight/display tracking, congurable
Output: Range
0/420 mA, congurable
Output: Resolution intern 16 bits = 65536 parts,
Resolution of 20,000 @ 20 mA
Output: Linearity error
@ 020 mA: <0.04%
@ 420 mA: 0.02%
Output: Temperature er-
ror
<100 ppm/K
Output: Zero point error 0.05%
Output: Max. error <0.1%
Output: Load
max. 0500
Output: Short-circuit
protected
yes
Output: Potential isolati-
on
yes
Inputs: 4× channel current or voltage input
020 mA, input impedance 250
010 V, input impedance 100 k
05 V, input impedance >10 M
Input: Resolution 3000 internal digits accordingly
e.g., 020 mA/010 V
Input: Accuracy 0.2%
Input: Linearity error <0.03%
Input: Temperature error <50 ppm/K
Input: Reserve
±15%, i.e., –1.5 V+11.5 V
Input: Potential isolation no
Cable type Screened twisted pair (e.g. LifYCY 2x2x0.20)
Cable length <150 m screened
Accessories D-Sub male connector, 15-pin
X3 Process Indicator PR 5410 4 Device installation
Minebea Intec EN-74

Table of Contents