Specications
Description Data
Connection D-Sub female connector, 15-pin
Output: Qty 1 active current output:
0/420 mA (max. 24 mA), 10 V output voltage via external
500 resistor
Output: Function Gross/net weight/display tracking, congurable
Output: Range
0/420 mA, congurable
Output: Resolution intern 16 bits = 65536 parts,
Resolution of 20,000 @ 20 mA
Output: Linearity error
@ 020 mA: <0.04%
@ 420 mA: 0.02%
Output: Temperature er-
ror
<100 ppm/K
Output: Zero point error 0.05%
Output: Max. error <0.1%
Output: Load
max. 0500
Output: Short-circuit
protected
yes
Output: Potential isolati-
on
yes
Inputs: 4× channel current or voltage input
020 mA, input impedance 250
010 V, input impedance 100 k
05 V, input impedance >10 M
Input: Resolution 3000 internal digits accordingly
e.g., 020 mA/010 V
Input: Accuracy 0.2%
Input: Linearity error <0.03%
Input: Temperature error <50 ppm/K
Input: Reserve
±15%, i.e., –1.5 V+11.5 V
Input: Potential isolation no
Cable type Screened twisted pair (e.g. LifYCY 2x2x0.20)
Cable length <150 m screened
Accessories D-Sub male connector, 15-pin
X3 Process Indicator PR 5410 4 Device installation
Minebea Intec EN-74

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish