EasyManua.ls Logo

Mobius AIRO Mobile CT System - E-Stop Test

Mobius AIRO Mobile CT System
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144 MI-42-0001 Rev. 19
MAINTENANCE: Service Functions
7.2.5 E-Stop Test
Overview An E-Stop, or Emergency Stop, test verifies that the scanner stops moving over the scan area and
ceases emitting x-rays in case of an emergency during a scan. Ring rotation does not stop
immediately. Always perform the test daily before setting up for procedures.
AIRO requires the E-Stop test to be run daily. To simplify the process, the system provides on-
screen, step-by-step instructions to walk operators and technicians through a “simulated use” test.
NOTE: In the event of an E-Stop failure discontinue use of the system and contact Technical
Support.
How to prepare for
a daily Emergency
Stop Test
Figure 117
Steps
1. Set up system for scanning. See sections 6.3 through 6.7.
2. Access
Hospital Service
. See sections 7.2.1 and 7.2.2.
3. Select E-Stop Test. See Figure 112.
4. Follow on-screen instructions.
5. Touch Next.
6. Set Start and End points for the scan.
NOTES:
Scan length minimum is 7.9cm (3.11 in,).
Set Start and End points for a scan length no longer than needed to press E-Stop
shortly after Z-axis translation starts and before the scan ends – approximately 10 to
15cm (4 to 6 in.).
Pressing the E-Stop button at the end of a long scan can result in a cool down wait
period.

Table of Contents