EasyManua.ls Logo

Mobius AIRO Mobile CT System - General Electromagnetic Immunity; Electromagnetic Emissions and Immunity

Mobius AIRO Mobile CT System
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MI-42-0001 Rev. 19 219
TECHNICAL DESCRIPTION: Regulatory Compliances and Approvals
9.12.2 General Electromagnetic Immunity
Electromagnetic
Environment
AIRO is intended for use in the electromagnetic environment specified in the following sections.
The owner of the equipment is responsible ensuring that the systems are used in such an
environment.
Use of this equipment adjacent to other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
Expected
Performance of ME
Equipment During
Immunity Testing
No unwanted motions or X-ray, no incorrect/corrupt transmission of data, and no effect on image
quality.
NOTE: The operator can expect that the transmission of data may be interrupted, but it will not be
lost. Data may have to be sent again.
Electromagnetic
Immunity
Declaration
The tables in the following sections provide guidance according to the manufacturer’s
electromagnetic immunity declaration.
9.12.3 Electromagnetic Emissions and Immunity
Emissions:
CISPR-11
IEC 61000-3-2
IEC 61000-3-3
Immunity:
IEC 61000-4-2,
IEC 61000-4-4,
IEC 61000-4-5,
IEC 61000-4-8,
IEC 61000-4-11,
IEC 61000-4-6,
IEC 61000-4-3
Emissions
T
est
Test Level and
Compliance Level
Electromagnetic Environment -
Guidance
Conducted and
Radiated RF Emissions
CISPR 11
Group 1, Class A
The AIRO is suitable for use in a typical
commercial or hospital environment.
NOTE: EMISSIONS characteristics of
this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11
class A). If it is used in a residential
environment (for which CISPR 11 class B
is normally required) this equipment might
not offer adequate protection to radio-
frequency communication services. The
user might need to take mitigation
measures, such as relocating or re-
orienting the equipment.
Harmonic Distortion
IEC 61000-3-2
Class A
Voltage Fluctuations
and Flickers
IEC 61000-3-3
Complies