TECHNICAL DESCRIPTION: X-ray Specifications
9.6 X-ray Specifications
9.6.1 Generator (High Voltage Power Supply)
Generator (HVPS)
Category Specification
Ty
pe High Frequency
Location Ring
Method of Cooling Air cooled
Power Rating [kW] 32
Peak power 32 kW
Duty Cycle
2 full energy scans per hour (120 kV x 250 mA x 20 sec, twice per
hour)
Input power Lithium Iron Phosphate LiFePO
4
battery system
Parameters Range
Rating
125 kV, 250 mA, 20 s
125 kV, 125 mA, 40 s
Output voltage 80 - 125 kV
Output current 5 - 250 mA
Nominal electric power (1s) 32 kW
Maximum average electric power 521 W
Minimum battery charge for proper operation 280 Vdc
Maximum Operating Points
120 kV, 250 mA, 21.75 s
120 kV, 125 mA, 43.5 s
30 minute cooling time between maximum
scans
Maximum Deviation from Indications
Tube Voltage: +/-1%
Tube current: +/- 2%
Tomogram scan times: User-defined
Topogram scan times: User-defined
Spiral scan times: User-defined