214 MI-42-0001 Rev. 19
TECHNICAL DESCRIPTION: Environmental Requirements
9.8 Environmental Requirements
Environmental
Requirements:
To avoid damaging the system in transport, storage, or unpacking, observe the
marked restrictions on the equipment and any protective packaging.
NOTE:
1. AIRO is installed per the installation procedure.
2. AIRO remains in original transportation packaging.
Least Favorable
Working Conditions
The least favorable working conditions are:
Least Favorable
Op
erational
Conditions
The least favorable operational conditions are:
Specification
Category
Operating and Hospital
Storage (see note 1)
Transport and Warehouse
Storage (see note 2)
Temperature 10°C (50°F) to 25°C (77°F) 0°C (32°F) to 50°C (122°F)
Humidity 20% to 70% non-condensing 5% to 95% condensing
Pressure 70 kPa to 110 kPa 70 kPa to 110 kPa
Local Restrictions
• Do not store near chemical products or gas.
• Do not expose to direct UV light.
• Protect from exposure to moisture, dust, and salinity.
Altitude <
3000 m (9840’) Not applicable
Adaptation Time
Minimum of twelve hours after
extreme storage conditions
Not applicable
Humidity
Preconditioning
Minimum of twelve hours after
extreme transport or storage
conditions
Not applicable
Parameter Least Favorable Condition
Temperature
At or above the high limit of the environmental temperature range:
25°C (77°F). Excessive temperature may degrade image quality.
Humidity
At or above the high limit of the environmental humidity range: 70%
non-condensing.
Atmospheric Pressure At or below the lowest limit of the atmospheric pressure: 70 kPa.
Parameter Condition
Scans per hour 2
Scan length 0.5 m
Tube loading 120 kV @ 250 mA