170 MI-42-0001 Rev. 19
MAINTENANCE: Preventative Maintenance
7.6 Preventative Maintenance
7.6.1 Preventative Maintenance is Required
Interval Preventative maintenance must be performed twice a year to ensure functionality and safety. The
owner/responsible organization is responsible for timely maintenance.
Exceeding the six month period for preventative maintenance is considered improper
use.
Authorization Only authorized service engineers or technicians are qualified to perform maintenance.
Risk of electrical shock. There are no user-serviceable parts. All servicing is to be
carried out by authorized service technicians.
Arrangement
Scope Preventative maintenance includes:
Complete inspection of the AIRO system
Potential adjustment of dosimetric accuracy
Adjustment or lubrication of normally wearing components
Replacement of system air filters
Service Contract Semi-annual Inspection
Yes
Automatically performed by authorized service engineers or
service technicians.
No
Contact technical support for arrangement. See section
1.2 Contact Data.