276 MI-42-0001 Rev. 19
DETACHABLE PENDANT: Technical Description for Detachable Pendants
13.4.2 Electromagnetic Emissions and Immunity with Detachable Pendant
Emissions and
Immunity:
IEC 61000-4-2
The item below shows supplemental information for Detachable Pendants on IEC 61000-4-2.
13.4.3 Location of Certification and Safety Labels
AIRO Certification
Labels on Systems
with Detachable
Pendants
NOTE: Systems with Detachable Pendants have labels as shown below.
Figure 188
Immunity
Test
IEC 60601 Test Level
and Compliance Level
Electromagnetic Environment - Guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 8 kV contact*
+/- 15 kV air*
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
* The Detachable Pendant has reduced levels of
+/- 4 kV contact and +/- 8 kV for multiple successive
shocks. It passes the higher levels for a single
shock.
No. Label in Location No. Label in Location
a
Columnless Warning (only on
columnless systems)
g
Support Contact (between Pendant
Holster and Ring)
s
Overbalancing Warning
h
Laser Warning
d
Emergency Stop button on Pendant
j
System Name Plate with CE, ETL, and
WEEE certifications
f
X-Ray Unit Warning
k