MI-42-0001 Rev. 19 211
TECHNICAL DESCRIPTION: X-ray Specifications
9.6.2 X-ray Source Assembly
X-ray Tube
NOTE: Accuracy of the Automatic Control System (ACS) is the same as that of the Tube Voltage
and current, +/- 2%.
9.6.3 Battery System
Battery System
Category Specification
Tu
be type Rotating Anode
Design Anode end grounded with metal housing
Beam shape Cone
Target 7 degrees rhenium-tungsten-molybdenum
Tube voltage and accuracy
80, 100, 120 kV
Accuracy: +/- 2%.
Tube current and accuracy
5 - 250 mA
Accuracy:
+/- 2%
Focal spot size 1.0 mm x 1.0 mm
Total filtration at max kV (inherent + beam
shaping filter) on central axis (mm Al
equivalent @ 120 kV)
6.8
Anode heat capacity (MHU) 1.7
Maximum anode cooling rate (kHU/min) 219.6
Method of tube cooling (FP-1000) 50% water, 50% propylene glycol
Method of tube cooling (MCS-1000) Oil - Shell Diala 34 ZX-1
Category Specification
Battery type LiFePO
4
Nominal terminal voltage 358 VDC
Total battery system energy storage 3580Wh
Battery Cell Rating Certificate UL1642 and IEC 62133
Battery Pack Rating Certificate UN/DOT 38.3 and IEC 62133