164 MI-42-0001 Rev. 19
MAINTENANCE: Cleaning, Disinfection, and Sterilization
7.4 Cleaning, Disinfection, and Sterilization
7.4.1 Cleanable Parts
Warnings and
Cautions
The CT system is not waterproof. It is not designed to protect the internal components from the
ingress of liquid.
Do not use sprays. Sprays may expose electrical components to moisture which can
damage the equipment.
Avoid using phenol-based, corrosive or solvent disinfection agents as they may harm
the surface material on the equipment.
Do not use high concentration (>500 ppm) sodium hypochlorite (bleach) solutions as
they may cause corrosion damage. Corrosion damage can lead to degradation of
structural elements that can result in patient/operator injury.
NOTE: If you are not sure of the properties of a cleaning or disinfectant agent, do not use it.
How to Clean Clean the system in an area that is acceptable for cleaning and disinfection of the AIRO, and that
contains enough room to perform this procedure. To clean the system, you will need the following:
Super Sani-Cloth Wipes (or equivalent)
Paper Towels
Bucket of warm water
Mild Detergent (dish washing soap or similar)
Cloth rags or wipes
Use caution when wiping down system as the batteries are still supplying power.
Steps
1. If not already done, unplug the AIRO from the wall mains.
2. Wear PPE as defined by your facilities bio-safety process.
3. Rotate the Gimbal to the Scan position.
See cautions about rotating the Ring in section
3.3.6 Rotating and Latching the Ring in Transport or Scan Position
4. Raise the Trumpf Column to its highest position to allow easier wipe down of the bellows.