MI-42-0001 Rev. 19 213
TECHNICAL DESCRIPTION: Performance Characteristics
9.7.2 Range, Accuracy, and Precision
Range and
Accuracy
Graph of Accuracy
of Radiation Output
Figure 161
Precision
Parameter Value
Rang
e 60 to 125 kVp
Accuracy +/- 10% over the range
Machine Settings
mA FS SDD
25 L 29”
Parameter Value
Pass/Fail Criteria 10%
Highest Percent Error 4.7% @ 60 kVp, 25 mA
Lowest Percent Error 1.2% @ 90 kVp, 25 mA

Table of Contents