MAINTENANCE: Cleaning, Disinfection, and Sterilization
NOTE: Only trained personnel should be performing this cleaning.
Ensure the Ring is tilted back to the vertical position prior to entering Transport Mode.
7.4.2 Sterilization
No Sterilization AIRO is not intended to be sterilized.
Disclaimers Mobius Imaging and distributors of the AIRO bear no
responsibility for sterilization of system
surfaces.
Mobius Imaging and distributors of the system shall not be held liable for any contagion which may
have stemmed from contaminated stains on the system surfaces.
4. Once in Hospital Service, press either the Up or Down button
s
, to tilt the Ring in the
required direction.
5. Once the tilt provides required access
d
, reseal the Tilt Stop mounting hole by carefully
screwing the Tilt Stop removed in step 2 back into position
f
.
To ensure that fluid and waste does not enter the system, always
reinstall Tilt Stops before cleaning the system
6. Clean and disinfect all Gimbal and Ring surfaces that become accessible when tilting the
Ring as outlined above.
NOTE: Fluid and waste can collect on the access panel at the bottom of the Gimbal. This
panel is only visible when the Ring is titled in Hospital Service mode. Always ensure that this
area of your AIRO is clean.
7. After the disinfection is complete, unscrew and retain the Tilt Stop reinstalled in step 5.
8. Return the Ring to the upright position.
9. Carefully reinstall the Tilt Stop removed in step 7, and tighten until snug.
Steps