MI-42-0001 Rev. 19 239
CT USER INFORMATION: Dosage Factors and Image Quality
Dose Profile: Body
Phantom
Figure 166
Variable Value
Head phantom position Center
120
kVp
100
mA
1.92
sec
192
mAs
Beam collimation
33.92
Exposure length Integral 119218.6
mrad-mm
Nominal slice thickness (mm) 33.92
CTDI (mrad) 3514.7
CTDI (mGy) 35.147
Variable Value
Body phantom position Center
120
kVp
100
mA
1.92
sec
192
mAs
Beam collimation
33.92
Exposure length Integral 43781.62 mrad-mm
Nominal slice thickness 33.92 mm
CTDI (mrad) 1290.73
CTDI (mGy) 12.91