32 MI-42-0001 Rev. 19
SAFETY: Mechanical Motion Safety
Avoiding Collisions
Do Not Lift There are no parts of AIRO that need to be lifted during authorized use with the exceptions of the
Pendant display and, for systems including a Trumpf TS7500 Operating Table Column, a remote
control for the Column and compatible tabletops.
Situation Risk Mitigation
Transferring the table
Evaluate the area for possible collisions.
Clear the area of objects, furniture, and bystanders before
any transfer.
Moving the Ring or the operating
table
Monitor all movements.
Stop movements by releasing the button on the controls.
Positioning tubing, other surgical
equipment
Anticipate the system path.
Place items so there is no danger of collision or pinching.
Changing table incline Monitor the changes in patient incline.
Moving the column to zero
position
Select Zero Position on the column only after evaluating
the area for possible collisions.
Pay particular attention to patient incline and any table
accessories that are angled downward.