7 Safety Evaluation Mode
Narda SRM-3006 97
7.2.3 Showing and hiding services
Safety Evaluation • Select Menu
You can specify the services that are to be measured in the Select Menu.
To show or hide specific services:
1. Press the Select Menu softkey.
ª The Select menu opens.
2. Select the services required using the rotary control and the softkeys.
You can find information on how to select items from a list under
Selecting entries from a list on page 47.
3. After you have selected the services, press the OK key.
ª The selected services will be displayed.
7.2.4 Setting the resolution bandwidth (RBW)
Safety Evaluation: RBW
Setting RBW automatically
The SRM-3006 sets the RBW automatically using the basic setting
Automatic:
This ensures that the narrowest service to be measured is resolved finely
enough.
RBW
Fspan narrowest service()
4
----------------------------------------------------------------------------
=