2.8 Rechargeable batteries
The SRM-3006 is fitted with a rechargeable Lithium-ion battery to allow
portable operation.
2.8.1 Storage
CAUTION
Unsuitable environmental conditions
Excessive temperatures and humidity can lead to a short circuit
which can result in a fire, which may cause injury or destroy the
battery pack.
Make sure that the storage conditions are within the following limits:
Temperature: -20 °C to +45 °C
Air humidity: 0% to 80%
Check regularly to ensure that the storage temperature and humidity
are within the prescribed limits.
Do not leave discharged batteries in the instrument for an extended
period of time.
Do not store the batteries for more than 6 months without recharging
them in the meantime.
Do not store the batteries together with any other metallic objects.
Long-term storage can reduce the battery capacity and shorten
battery life. The casing can also be damaged internally by leaking
electrolyte.