12 Spatial Averaging
Narda SRM-3006 145
The averaging process
The SRM-3006 simplifies the measurement by automatically averaging the
values measured at different points in the room. The SRM-3006 determines
the root mean square value (RMS), i.e. the average power level. For this
reason it is a good idea to set the detector function in Level Recorder mode
to RMS, although the PEAK setting is also available.
Spatial Averaging can be applied in the following operating modes:
Safety Evaluation (see page 93)
Spectrum Analysis (see page 105)
Level Recorder (see page 135)
It cannot be used in UMTS P-CPICH Demodulation mode.
According to the operating mode, the SRM-3006 determines the average of:
the individual results for the different services (Safety Evaluation mode),
the individual spectral values (Spectrum Analysis mode), or
the individual measured values (Level Recorder mode).
12.2 Description of averaging functions
The following types of Spatial Averaging are available:
Continuous: measured values are recorded continuously
Discrete: one measured value recorded each time a key is pressed
Discrete Axis: one single axis measured value recorded each time a key
is pressed
These three averaging functions are described briefly below.