17 Specifications
Narda SRM-3006 201
17.6 Single axis E-field antenna 3531/04
17.6.1 Characteristics
1 Typical measurement dynamic range for 10 dB signal to noise radio (RBW = 1 kHz)
2 Typical values
17.6.2 Uncertainty
1 Typical value k = 2 (K= extrapolation or correction factor for determining the assessment value);
+15 °C to +30 °C
17.6.3 General specifications
Frequency range 9 kHz – 300 MHz
The correction factors determined individually
during calibration are stored in an EEPROM and
are applied automatically when used in conjunction
with the SRM-3006 basic unit.
Antenna type E-field
Sensor type Single axis active broadband dipole
Dynamic range
1)
• 70 µV/m – 16 V/m for 300 kHz – 10 MHz
• 70 µV/m – 36 V/m for > 10 MHz – 300 MHz
CW damage level > 1000 V/m
Intrinsic noise display in
conjunction with the SRM-3006
basic unit
2)
20 µV/m in the range from 1 MHz – 300 MHz with
1 kHz resolution bandwidth (RBW)
Measurement range limit (for
single CW signal)
2)
50 V/m
RF connector N connector, 50 Ω
Extended measurement
uncertainty
1)
(in conjunction with SRM-3006
basic unit and 1.5 m RF cable)
Frequency range Single axis
measurement
0.1 – 300 MHz 2.0 dB
Calibration uncertainty < 1.2 dB
Operating temperature range -10 °C to +50 °C (same as SRM-3006 basic unit)
Air humidity < 29 g/m
3
(< 93 % at +30 °C)