17 Specifications
Narda SRM-3006 195
17.3 Three axis E-field antenna 3502/01
17.3.1 Characteristics
1 Typical measurement dynamic range for 10 dB signal to noise ratio (RBW = 1 kHz);
1.8 to 2.2 GHz
2 Typical values
Frequency range 420 MHz to 6 GHz
The correction factors determined individually
during calibration are stored in an EEPROM and
are applied automatically when used in conjunction
with the SRM-3006 basic unit.
Antenna type E-field
Sensor type Three axis design with scanned axes
Dynamic range
1)
0.14 mV/m to 160 V/m
CW damage level 435 V/m or 50 mW/cm
2
Intrinsic noise display in
conjunction with the SRM-3006
basic unit (separate
measurement of a single axis)
2)
33 µV/m at 900 MHz with 1 kHz resolution
bandwidth (RBW)
25 µV/m at 2.1 GHz with 1 kHz resolution
bandwidth (RBW)
Intrinsic noise display in
conjunction with the SRM-3006
basic unit (for isotropic result)
2)
60 µV/m at 900 MHz with 1 kHz resolution
bandwidth (RBW)
43 µV/m at 2.1 GHz with 1 kHz resolution
bandwidth (RBW)
Measurement range limit
(for single CW signal)
200 V/m
Max. measurement range (in
conjunction with the SRM-3006
basic unit)
2)
160 V/m (without restrictions for total span of
750 MHz to 6 GHz)
RF connector N connector, 50 Ω