Narda SRM-3006 9
1.3.1 Basic unit
The SRM-3006 Basic Unit has the following standard features of a spectrum
analyzer:
•The Integration over Frequency Band function in Spectrum Analysis
mode can be used to determine the wide band value for an individual
channel (Channel Power).
• The service tables in Safety Evaluation mode can be used as channel
tables for channel measurements (Channel Power Plus). Clarity is given
by displaying the results as a bar graph.
• Noisy signals can be smoothed by adjusting the video bandwidth or by
selecting a narrow RBW in Safety Evaluation mode. A compromise with
regard to time must be found when smoothing the signals.
• Signals can be observed against time in Spectrum Analysis mode using
very narrow and very wide RBWs.
In addition to the above, the SRM-3006 has the following functions (among
others):
• The field strength of pulsed signals can be determined rapidly in Level
Recorder mode.
• Signals can be analyzed in real time or monitored over a long period of up
to 24 hours in Scope mode.
• Spatial averaging and normal averaging provide important information for
standard-compliant measurements e.g. for 6-minute intervals.
• Peak Table and Extrapolation functions are provided for evaluating
measurements.
• Demodulation enables UMTS mode.
• Certain store events (time, threshold value) can be set to enable recording
of specific measurement results.
1.4 PC software
The device can be connected to a PC via the optical interface or the USB
port (Mini B type). This allows access to the device functions and
measurement results.
• Configuration of the measuring set using the SRM-3006 Tools.
• Remote control commands (see Remote Control on page 175)