17 Specifications
Narda SRM-3006 193
17.2.2 Uncertainty
1 Typical value k = 2 (k = extrapolation or correction factor for calculating the assessment value);
+15 °C to +30 °C
17.2.3 General specifications
Compliance
Extended
measurement
uncertainty
1)
(in conjunction with
SRM-3006 basic
unit and 1.5 m RF
cable)
Frequency Range Single axis
measurement with
isotropic antenna
Isotropic
measurement
27 – 85 MHz +2.4 / -3.3 dB +3.2 / -4.8 dB
> 85 – 900 MHz +2.4 / -3.4 dB +2.5 / -3.6 dB
> 900 – 1400 MHz +2.3 / -3.1 dB +2.5 / -3.4 dB
> 1400 – 1600 MHz +2.3 / -3.1 dB +2.6 / -3.8 dB
> 1600 – 1800 MHz +1.8 / -2.3 dB +2.2 / -3.0 dB
> 1800 – 2200 MHz +1.8 / -2.3 dB +2.4 / -3.3 dB
> 2200 – 2700 MHz +1.9 / -2.4 dB +2.7 / -3.8 dB
> 2700 – 3000 MHz +1.9 / -2.4 dB +3.3 / -5.3 dB
Calibration
uncertainty
< 1.5 dB
Operating temperature range -10 °C to +50 °C (same as SRM-3006 basic unit)
Air humidity < 29 g/m
3
(< 93 % at +30 °C)
RF immunity 200 V/m between 27 MHz and 3 GHz
Climatic Storage 1K3 (IEC 60721-3),
extended to-10 °C to +50 °C
Transport 2K4 (IEC 60721-3)
Operating 7K2 (IEC 60721-3)
Mechanical Storage 1M3 (IEC 60721-3)
Transport 2M3 (IEC 60721-3)
Operating 7M3 (IEC 60721-3)
ESD und EMC EN 61326:2006
Safety EN 61010-1:2004
EU Guidelines 2003/11/EG 06.02.2003 (PBDE and OBDE)
2002/95/EG 27.01.2003 (RoHS)
2002/96/EG 27.01.2003 (WEEE)
CE (European Community) Ye s ( Declaration of Conformity on page 205)