17 Specifications
196 SRM-3006 Narda
17.3.2 Uncertainty
1 Typical value k = 2 (k = extrapolation or correction factor for calculating the assessment value
+15 °C to +30 °C
17.3.3 General specifications
Compliance
Extended
measurement
uncertainty
1)
(in conjunction with
SRM-3006 basic
unit and 1.5 m RF
cable)
Frequency range Single axis
measurement with
isotropic antenna
Isotropic
measurement
420 – 750 MHz +2.1 / -2.9 dB +2.6 / -3.8 dB
> 750 – 1600 MHz +2.0 / -2.7 dB +2.2 / -2.9 dB
> 1600 – 2000 MHz +1.7 / -2.2 dB +1.9 / -2.4 dB
> 2000 – 4000 MHz +1.7 / -2.2 dB +2.0 / -2.6 dB
> 4000 – 4500 MHz +1.8 / -2.3 dB +2.2 / -3.0 dB
> 4500 – 5000 MHz +1.9 / -2.5 dB +2.5 / -3.5 dB
> 5000 – 6000 MHz +1.9 / -2.5 dB +2.9 / -4.3 dB
Calibration
uncertainty
< 1.5 dB
Operating temperature range -10 °C to +50 °C (same as SRM-3006 basic unit)
Air humidity < 29 g/m
3
(< 93 % at +30 °C)
RF immunity 200 V/m
Climatic Storage 1K3 (IEC 60721-3),
extended to -10 °C to +50 °C
Transport 2K4 (IEC 60721-3)
Operation 7K2 (IEC 60721-3)
Mechanical Storage 1M3 (IEC 60721-3)
Transport 2M3 (IEC 60721-3)
Operation 7M3 (IEC 60721-3)
ESD and EMC EN 61326:2006
Safety EN 61010-1:2004
EU Guidelines • 2003/11/EG 06.02.2003 (PBDE and OBDE)
• 2002/95/EG 27.01.2003 (RoHS)
• 2002/96/EG 27.01.2003 (WEEE)
CE (European Community) Ye s ( Declaration of Conformity on page 205)
Country of origin Germany