EasyManua.ls Logo

NARDA SRM-3006 - Measuring in the Gaps between Defined Services (Others); Using the Table View

NARDA SRM-3006
240 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Safety Evaluation Mode
Narda SRM-3006 99
7.2.7 Measuring in the gaps between defined services
(Others)
Safety Evaluation • Others
The field strengths of the services not selected in the service table can be
measured and their overall value shown by activating the Others function.
Note:
Only services that are not in the service table but within the frequency
range of the hardware (antenna / cable) are taken into account.
To activate and deactivate the function:
Press the Others key until the desired status (On, Off) is displayed.
ª The table view shows Others; the graphical view shows O.
Tip: The measurement can be speeded up by deactivating Others.
7.3 Using the Table view
The services and measured values are displayed as a table in the Table
view. The values displayed depend on the table type (see page 100) and
from the selected result types (see page 75).
Figure 20: Table view in Safety Evaluation mode
Note: Information about the status bars is found under The upper status
bar on page 43 and The lower status bar on page 44.

Table of Contents

Other manuals for NARDA SRM-3006

Related product manuals