6 Functions Common to More Than One Operating Mode
Narda SRM-3006 73
6.3.1 Manual measurement range selection
Manual adjustment is recommended if the field situation is known or if
pulsed signals are being measured.
1. The Set Meas. Range selection box opens when you press the Meas.
Range softkey. It displays a list of possible input sensitivities.
Figure 16: Set Measurement Range selection box
2. Use the rotary control to select the desired input sensitivity.
3. Press the OK key.
ª The input sensitivity is set.
Note: The selected measurement range applies to all operating modes and
does not have to be set again if you change operating mode.